Details
The CHAD Evolution Electronic Oygen Conserver has a minimum of a 2-year battery life.
Additional Information
| Features |
- Delivers 5:1 savings ratio at all settings.
- Minimum 2 year battery life utilizing 2 AA Batteries (average 4 hrs per day usage).
- Delivers a uniform oxygen pulse from 14-40 breaths per minute.
- Liter flow increments: 1, 2, 3, 4, 5, 6, 7 Plus CF preset @ 2 LPM.
- Operating pressure range: 200-3000 PSI.
- Senses a breath/delivers oxygen in the first half of the inspiratory cycle.
- Single push button selects the equivalent LPM dosage settings.
- Weighs only 14.9 ounces with batteries.
- Auto shutoff feature conserves battery life.
- Fits easily in virtually any carry bag.
- 2 year limited warranty.
|
| Items Included |
No |
| Specifications |
| Setting | | Cylinder Type | Cylinder Volume | 1 | 2 | 3 | 4 | 5 | 6 | 7 | CF 2 lpm | | Estimated Cylider Duration in Hours (Based on 20 Breathes per Minute) | | M2 | 36 Liters | 3.0 | 1.5 | 1.0 | 0.8 | 0.6 | 0.5 | 0.4 | 0.3 | | M4 (A) | 113 Liters | 9.4 | 4.7 | 3.1 | 2.4 | 1.9 | 1.6 | 0.7 | 0.9 | | M6 (B) | 164 Liters | 13.7 | 6.8 | 4.6 | 3.4 | 2.7 | 2.3 | 1.9 | 1.4 | | ML6 | 171 Liters | 14.3 | 7.1 | 4.8 | 3.6 | 2.9 | 2.4 | 2.0 | 1.4 | | M7 | 198 Liters | 16.5 | 8.3 | 5.5 | 4.1 | 3.3 | 2.8 | 2.3 | 1.7 | | M9 (C) | 246 Liters | 20.5 | 10.3 | 6.8 | 5.1 | 4.1 | 3.4 | 2.9 | 2.1 | | D | 425 Liters | 35.4 | 17.7 | 11.8 | 8.9 | 7.1 | 5.9 | 5.0 | 3.5 | | E | 680 Liters | 56.7 | 28.3 | 18.9 | 14.2 | 11.3 | 9.4 | 8.1 | 5.7 |
|